Page 215 - Praxair - Specialty Gases and Equipment Reference Guide
P. 215

Laboratories

            Analytical Instrumentation





              Praxair’s Gas Grade Recommendations
              Analytical Method/     Instrument   Page     Analytical Range     Type of Analysis
              Detector               Gases        Number   < 100 ppm   > 100 ppm  Impurity Considerations
              Gas Chromatography
              TCD                    Ar           B •26    AR 5.5TG   AR 5.0UH  Universal Detector
              (Thermal Conductivity Detector)   He   B •53   HE 5.5TG   HE 5.0UH  Atmospheric contaminants can oxidize
                                                                                the detector filament giving rise to
                                     H 2          B •54    HY 6.0RS   HY 5.0UH  negative peaks and reduced sensitivity.
                                     N 2          B •65    NI 5.5TG   NI 5.0UH
              FID                    Air          B •24    AI 0.0UZ   AI 0.0UZ  Organic Compounds
              (Flame Ionization Detector)   Ar    B •26    AR 5.5TG   AR 5.0UH/  Hydrocarbons in carrier and fuel gases
                                                                     AR 4.8Z    can give rise to baseline noise and
                                     He           B •53    HE 5.5TG   HE 5.0UH/  reduced detector sensitivity. Oxygen
                                                                                and water cause column deterioration
                                                                     HE 4.6Z    and affect retention time on critical
                                     H 2          B •54    HY 6.0RS   HY 5.0UH/  separations.
                                                                     HY 4.5Z
                                     N 2          B •65    NI 5.5TG   NI 5.0UH/
                                                                     NI 4.8Z
                                     40% H 2  in He   D •218   IG F12UH   IG F11
                                     40% H 2  in N 2    D •218   IG F14UH   IG F13
              ECD                    He           B•53     HE 5.5EC/   HE 5.5TG  Electronegative Functional Groups
              (Electron Capture Detector)                  HE 5.5TG             Detector response and column life are
                                     N 2          B•65     NI 5.5EC/   NI 5.5TG      reduced by oxygen and water. Hydro-
                                                           NI 5.5TG             carbons and halocarbons can produce
                                     5% CH 4  in Ar   D •218   IG ECD1   IG ECD1      baseline noise, negative peaks, and
                                                                                plumbing contamination.
     Laboratories
                                     (ECD P-5)
                                     10% CH 4  in Ar   D •218   IG ECD2   IG ECD2
                                     (ECD P-10)
              FPD                    Air          B •24    AI 0.0UZ   AI 0.0Z   Sulfur or Phosphorous Compounds
              (Flame Photometric Detector)   He   B •53    HE 5.5TG   HE 5.0UH  Organics can yield baseline noise and
                                                                                carbon dioxide can suppress detector
                                     H 2          B •54    HY 6.0RS   HY 5.0UH  response.
                                     N 2          B •65    NI 5.5TG   NI 5.0UH
              PID                    He           B •53    HE 5.5TG   N/A       Selective Detector Dependent
              (Photo Ionization Detector)   H 2    B •54   HY 6.0RS   N/A       on UV Source
                                                                                Organics can yield baseline noise and
                                     N 2          B •65    NI 5.5TG   N/A       carbon dioxide can suppress detector
                                                                                response.
              MS                     Ar           B •26    AR 5.5TG   AR 5.0UH  Universal Detector
              (Mass Spectrometer)    He           B •53    HE 5.5TG   HE 5.0UH  Organics can yield baseline noise and
                                                                                carbon dioxide can suppress detector
                                     H 2          B •54    HY 6.0RS   HY 5.0UH  response.
                                     N 2          B •65    NI 5.5TG   NI 5.0UH
              DID                    He           B •53    HE 6.0RS   N/A       Universal Detector
              (Discharge Ionization Detector)   He Purge   B •53   HE 5.0UH   N/A   Atmospheric impurities can cause base-
                                                                                line noise signal polarity and reduced
                                                                                detector stability and sensitivity.
              USD                    Ar           B •26    AR 5.0UH   AR 5.0UH/  Universal Detector
              (Ultrasonic Detector)                                  lAR 4.8Z   Atmospheric impurities can cause base-
                                     He           B •53    HE 5.0UH   HE 5.0UH/  line noise signal polarity and reduced
                                                                     HE 4.6Z    detector stability and sensitivity.